Page 14 - PJ Product e-Book Manual
P. 14
size 421mm(L)*570mm(W)*435mm(H)
Parameter Range
Injection Mode Full cycle injection/partial injection/microliter carrier injection
Quantitative ring volume Standard 20mL, optional 5mL/10mL/25mL (customizable)
Self developed core
Integrated cleaning station flow path components syringe pump Standard 25mL, optional 10mL/25mL, etc
External cleaning solvent bottle, cleaning before and after injection,
multiple cleaning agents to choose from,
Wash needles It has a drying function and can choose whether to clean between
samples or each needle
Full cycle injection: RSD 0.3%
Sample repeatability Partial injection: RSD 0.5%
Non destructive injection: RSD 0.5%
Sample capacity 132 15ml sample tubes, etc
Two sample trays can be selected freely
(supporting customized sample tube specifications)
cross-contamination <0.1%
linear >0.999
Maximum withstand voltage 35MPa (customizable for higher pressure requirements)
Beautiful appearance and small size
�� ��